Title:
Dynamic Pressure Absolute Calibration Method
Author(s):
KongDeren SunHaibo ZhuMingwu LiYongxin
School of Mechanics,NUST,Nanjing 210094
Keywords:
pressure measurement error analysis precision metrolog y unit
PACS:
TB935
DOI:
-
Abstract:
T his paper describes drop hammer hydraulic dynamic-calibrat ion device. s operating principle and it s main funct ion. The problems of the current calibration methods are also studied in details. In order to improve the calibrat ion system’s accuracy, an absolute calibration method is introduced. A low-price and high-accuracy force transducer is used to replace pressure transducer to realize dynamic pressure absolute calibrat ion. The significance of the method is analy zed. The way of how to put the method into effect , and the theoret ical proof of the calibration accuracy are all stated. The experiments have proved the feasibility of the method.

References:

1 朱明武, 王宗支, 梁人杰等. 动压测量. 北京: 国防工业出版社, 1984
2 孔德仁. 压力动态绝对校准方法研究. 南京: [ 硕士学位论文] . 南京: 南京理工大学, 1997